10 Technical data
General specifications
Manufacturer Endress+Hauser
Product designation Liquisys M CPM 223
Liquisys M CPM 253
Input
Measured variables pH, redox, temperature
pH measurement
Measuring range pH –2 ... 16
Display range pH –2 ... 16
with indication of measuring range violation
pH offset range ±pH 2
Slope Glass: 38.00 ... 65.00 mV/pH (nominal 59.16 mV/pH)
Antimon: 25.00 ... 65.00 mV/pH (nominal 59.16 mV/pH)
IsFET: 38.00 ... 65.00 mV/pH (nominal 59.16 mV/pH)
Zero point Glass: pH 5.00 ... 9.00 (nominal pH 7.00)
Antimon: pH –1.00 ... 3.00 (nominal pH 1.00)
IsFET: –500 ... +500 mV
Terminal cross-section 2.5 mm
2
Required conductor cross-section CPM 223
CPM 253
0.75 mm
2
/ AWG 18
1.50 mm
2
/ AWG 14
Maximum cable length to pH electrode 50 m
pH signal input
Input resistance for nominal operating conditions > 1 × 10
12
Ω
Redox measurement
Display and measuring range –1500 ... +1500 mV / 0 ... 100 %
Redox offset range ±120 mV / ±50 %
Terminal cross-section 2.5 mm
2
Required conductor cross-section CPM 223
CPM 253
0.75 mm
2
/ AWG 18
1.50 mm
2
/ AWG 14
Maximum cable length to pH electrode 50 m
Redox signal input
Input resistance for nominal operating conditions > 1 × 10
12
Ω
Temperature measurement
Temperature sensor Pt 100, Pt 1000, NTC 30K
Measuring range Pt 100, Pt 1000
NTC 30K
–50 ... +150 °C
–20 ... +100 °C
Temperature offset range ±5 °C
Digital inputs 1 and 2
Voltage 10 ... 50 V
Current consumption max. 10 mA
Current input
Current range 4 ... 20 mA, galvanically separated
Load 260 Ω at 20 mA (voltage drop 5.2 V)
%
Ω
°C
PM253E10.CHP
Liquisys M CPM 223 / 253 Technical data
Endress+Hauser 79