Note:
The instrument is pre-configured
for symmetrical measurement.
Change the configuration in field
A2 for asymmetrical measurement
(see chapter 5.2.1).
The conductor for the potential
matching pin must be connected
to the PA/PMterminal of the
instrument for symmetrical
measurement.
If the software setting
asymmetricalis chosen for a
symmetrical connection, this will
reduce the service life of the
reference electrode.
The instrument is approved for
protection class II and is generally
connected without protective earth.
Cable termination is only required
for CPM 223.
PH-AN01E.CDRFig. 3.18
pH electrode connection
to Liquisys M
CPM 223 / 253
left:
Connection of CPS 11
with CPK 9
right:
Connection of CPS 401
with CPK12
RE-AN01E.CDR
Fig. 3.19
Redox electrode
connection to Liquisys M
CPM 223 / 253
left:
Asymmetrical (without PM)
right:
Symmetrical (with PM)
Installation Liquisys M CPM 223 / 253
16 Endress+Hauser