Control characteristic for direct and inverted control action
Field R236 offers two control characteristics for selection which have the effects shown in the
following diagram.
5.6.3 Timer for cleaning function
This function can be used to implement a
simple cleaning routine. The user can specify
a time interval after which cleaning is to start;
i.e. only constant intervals can be defined.
More extended cleaning functions can be
implemented in conjunction with the
ChemoClean function (version with four
contacts, see chapter 5.5.4).
Note:
The timer and ChemoClean do not
work independently of each other.
Whilst one of the functions is active,
the other cannot be started.
X
X
– X
– X
w
w
w
100 %
100 %
Y
Y
direct =
max. function
inverted =
min. function
REGEL03.CDR
Fig. 5.13
Control characteristic of a
proportional-action
controller with direct and
inverted control action
Wiper and / or spray cleaning
system
active
inactive
Hold function
active
inactive
Cleaning time
(0 ... 999 s)
Pause time between
2 cleaning intervals
(1 ... 7200 min)
Clean-Hold
dwell period
(0 ... 999 s)
Time
DIAGR-TI.CDR
Fig. 5.14
Relationship among
cleaning time, pause time
and hold dwell period
PM253E5B.CHP
Liquisys M CPM 223 / 253 Instrument configuration
Endress+Hauser 41