3.4 Electrical connection
Connection diagram
The connection diagram depicted in Fig. 3.10 shows the connections of an instrument
equipped with all the options. The connection of the various electrodes with the measuring
cables is shown in more detail in the subsequent figures.
Note:
• The instrument has protection
class II and is generally operated
without protective earth
connection.
• Mains supply voltage fluctuations
should not exceed ten percent of
the nominal supply voltage.
• 24V AC/DC models must be
supplied from an energy limiting
SELV source in accordance with
dir. IEC 1010.1 Annex H.
• Ground the sensor screen in order
to ensure functional safety and
measuring stability of the
measuring system.
Glass electrodes: Terminal “S”.
IsFET: PE ground terminal. The
ground terminal is located on the
cover frame of the panel-mounted
instrument CPM 223 and in the
connection compartment in the
field instrument CPM 253.
pH/SRC
Ref.
11
12
13
47
48
49
57
58
59
51
52
53
54
55
56
PA/
PM
15 V
85
86
S
10-50 V
81
93
82
94
10-50 V
L+
L-
L1
N
23 +
24 -
mA
∼=
41
42
43
pH
mA
31
32
mA
33
34
Aux. voltage
output
Digital
input 2
(Chemoclean)
Digital
input 1
(Hold)
Signal output 2
temperature
Signal output 1
pH / redox
Temperature
Electrode
Alarm
(current-free
contact
position)
Relay 1
(current-free
contact
position
Relay 2
(current-free
contact
position)
Relay 3
(current-free
contact
position)
Relay 4
(current-free
contact
position)
Mains
GN
WH
YE
ANPH01.CDR
Ground glass electrodes
Ground IsFET
Current input
4...20 mA
Fig. 3.10
Electrical connection of
Liquisys M CPM 223/253
with full wiring
Installation Liquisys M CPM 223 / 253
12 Endress+Hauser